Securitas Healthcare LLC: Device Recall

Recall #Z-1032-2026 · 12/08/2025

Class II: Risk

Recall Details

Recall Number
Z-1032-2026
Classification
Class II
Product Type
Device
Recalling Firm
Securitas Healthcare LLC
Status
Ongoing
Date Initiated
12/08/2025
Location
Lincoln, NE, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1898 units

Reason for Recall

Securitas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing.

Product Description

Arial 900 MHz Call Station, Model Number CSK200-900MR, with Firmware version: 1.6.3

Distribution Pattern

Worldwide distribution -US Nationwide and the country of Canada.