ICU Medical, Inc.: Device Recall
Recall #Z-0965-2026 · 11/24/2025
Class II: Risk
Recall Details
- Recall Number
- Z-0965-2026
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- ICU Medical, Inc.
- Status
- Ongoing
- Date Initiated
- 11/24/2025
- Location
- Lake Forest, IL, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 7 customers
Reason for Recall
For certain customers, the Change Summary review option ceases to operate as expected for any migrated library after the migration has been completed. Users must use alternate methods to review changes. Failure to do so may lead in over delivery, under delivery, or delay in setup on the infusion pump.
Product Description
LifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02
Distribution Pattern
US Nationwide distribution in the states of CA, IA, LA, NY, PA, TN, VA.