Intuitive Surgical, Inc.: Device Recall
Recall #Z-0964-2026 · 08/21/2024
Class II: Risk
Recall Details
- Recall Number
- Z-0964-2026
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Intuitive Surgical, Inc.
- Status
- Ongoing
- Date Initiated
- 08/21/2024
- Location
- Sunnyvale, CA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 144
Reason for Recall
Surgical system Error 23062 due to connector1) Partially damaged "Potential problem with ergonomics: Disable - Retry " or 2) Failed "Disable Ergonomics: Viewer, headrest, armrest, and pedal tray position cannot be adjusted until da Vinci is restarted. Contact Technical Support to report the issue: Cancel-Disable" ergonomics locked until motor replaced; may lead to delay/aborted/alternate procedure
Product Description
da Vinci 5, Surgeon Console Column Motor Connector; ASSY,DV5 CONSOLE,SSC,IS5000; Part Number: 380730-40
Distribution Pattern
US Nationwide distribution in the states of MN, FL, CO, CA, NC, KS, TX, SC, WV, LA, OH, NY, NH, GA, KY, OK, NJ, MA, OR, PA, NM, AZ, TN, ND, IL, AR, WA, IN, VA, CT.
Other Recalls by Intuitive Surgical, Inc.
- Class II: Risk 12/15/2025
- Class II: Risk 12/04/2025
- Class II: Risk 11/07/2025
- Class II: Risk 02/26/2025
- Class II: Risk 12/19/2024
Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.