Percussionaire Corporation: Device Recall

Recall #Z-0960-2025 · 12/23/2024

Class I: Dangerous

Recall Details

Recall Number
Z-0960-2025
Classification
Class I
Product Type
Device
Recalling Firm
Percussionaire Corporation
Status
Ongoing
Date Initiated
12/23/2024
Location
Sandpoint, ID, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4,727 packs/cases

Reason for Recall

Due to internal venturi component getting transiently stuck in the forward position within the breathing circuit and may cause a reduction of amplitude and volume of ventilation.

Product Description

VDR4 Phasitron Breathing Circuit Models that contain the venturi component: -A50094 D 5PK, Phasitron Kit, VDR, Single Patient, 5pk -A51001, VDR4 Control Unit Tester -PRT 991, VDR W/SWIVEL T SINGLE PATIENT PHASITRON -PRT 992, VDR4 HUMIDIFICATION ADAPTER KIT, CASE OF 10 -PRT 993, VDR4 Humidification Kit with Cross Tee, Case of 10 The Phasitron breathing circuit kit is intended to be used for continuous ventilation intended for controlled ventilation of patients

Distribution Pattern

Worldwide - U.S. Nationwide distribution in the states of AZ, FL, IL, MS, MO, NE, NJ, NY, NC, RI, TX, UT, and WV. The countries of Belgium, Canada, France, Hong Kong, Netherlands, Russia, Switzerland and Turkey.