Percussionaire Corporation: Device Recall
Recall #Z-0960-2025 · 12/23/2024
Recall Details
- Recall Number
- Z-0960-2025
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- Percussionaire Corporation
- Status
- Ongoing
- Date Initiated
- 12/23/2024
- Location
- Sandpoint, ID, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4,727 packs/cases
Reason for Recall
Due to internal venturi component getting transiently stuck in the forward position within the breathing circuit and may cause a reduction of amplitude and volume of ventilation.
Product Description
VDR4 Phasitron Breathing Circuit Models that contain the venturi component: -A50094 D 5PK, Phasitron Kit, VDR, Single Patient, 5pk -A51001, VDR4 Control Unit Tester -PRT 991, VDR W/SWIVEL T SINGLE PATIENT PHASITRON -PRT 992, VDR4 HUMIDIFICATION ADAPTER KIT, CASE OF 10 -PRT 993, VDR4 Humidification Kit with Cross Tee, Case of 10 The Phasitron breathing circuit kit is intended to be used for continuous ventilation intended for controlled ventilation of patients
Distribution Pattern
Worldwide - U.S. Nationwide distribution in the states of AZ, FL, IL, MS, MO, NE, NJ, NY, NC, RI, TX, UT, and WV. The countries of Belgium, Canada, France, Hong Kong, Netherlands, Russia, Switzerland and Turkey.