Cepheid: Device Recall

Recall #Z-0959-2026 · 10/30/2025

Class II: Risk

Recall Details

Recall Number
Z-0959-2026
Classification
Class II
Product Type
Device
Recalling Firm
Cepheid
Status
Ongoing
Date Initiated
10/30/2025
Location
Sunnyvale, CA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6325 units

Reason for Recall

Invitro diagnostic test kits with specimens that have elevated white blood cell count (>30,000 cell) may cause inaccurate results.

Product Description

Brand Name: Xpert¿ BCR-ABL Ultra Model/Catalog Number: GXBCRABL-US-10

Distribution Pattern

US: AL AR AZ CA CO FL GA IA IL KS KY LA MD ME MI MN MO NC NE NH NJ NY OH PA RI TX VA WA WI OUS: Algeria Armenia Australia Austria Bahrain Bangladesh Belgium Brazil Burkina Faso Cambodia Canada Canary islands Chile Colombia Cyprus Czech Republic Dem. Rep. Congo Djibouti Dominican Rep. Ecuador El Salvador Eswatini Finland France Frenc.Polynesia Gambia Georgia Germany Ghana Greece Guadeloupe Guatemala Honduras Hong Kong India Indonesia Iraq Israel Italy Jamaica Jordan Kazakhstan Kenya Kuwait Latvia Lebanon Lithuania Luxembourg Malawi Malaysia Mauritania Mauritius Mexico Morocco Mozambique Myanmar Netherlands New Caledonia New Zealand Nicaragua Norway Oman Pap. New Guinea Peru Philippines Poland Portugal Qatar Rep.of Congo Reunion Rwanda Saudi Arabia Senegal Seychelles Singapore South Africa Spain Sri Lanka Sweden Switzerland Taiwan Tajikistan Tanzania Thailand Togo Tunisia Uganda United Kingdom Uruguay Utd.Arab Emir. Venezuela Zambia Zimbabwe

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.