SIN SISTEMA DE IMPLANTE NACIONAL S.A: Device Recall

Recall #Z-0920-2025 · 09/01/2024

Class II: Risk

Recall Details

Recall Number
Z-0920-2025
Classification
Class II
Product Type
Device
Recalling Firm
SIN SISTEMA DE IMPLANTE NACIONAL S.A
Status
Ongoing
Date Initiated
09/01/2024
Location
Sao Paulo, Brazil
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
123 (US 98, OUS 25)

Reason for Recall

Dental implant box incorrectly labels the diameter as 4,5 X 18,0 mm, but the internal label correctly labels the diameter as 4,0 X 18,0 mm; if the dentist drills the bone with the larger diameter stated on the box, the smaller diameter implant, will become loose and will not fit, which may necessitate an additional surgery.

Product Description

Implant Epikut S Plus CM 16-Degrees 4,5 X 18,0 mm, REF: ILM 4518N

Distribution Pattern

US distribution to states of: KY, CO, CA, NY, ID, NJ, FL, TX, KY, IL, OH; AND OUS (foreign) country of Brazil.