GE Medical Systems China Co., Ltd.: Device Recall
Recall #Z-0915-2026 · 11/14/2025
Recall Details
- Recall Number
- Z-0915-2026
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- GE Medical Systems China Co., Ltd.
- Status
- Ongoing
- Date Initiated
- 11/14/2025
- Location
- Wuxi, China
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 203 units
Reason for Recall
Potential unexpected shutdown of Carestation 600 and 700 Series Anesthesia Systems containing certain power management boards, if the AC mains power is unplugged or in the event of an AC mains power failure. Anesthesia systems only operate on battery power in a rare event that AC mains power is lost and there is no continuous backup emergency power. If AC power is interrupted, the system will not automatically switch over to the battery supply mode and will reboot. If this issue occurs, a temporary disruption of mechanical ventilation, manual ventilation, and volatile agent delivery may occur. Following the reboot, the system will not return to the previous ventilation settings.
Product Description
GE Healthcare Carestation anesthesia system Power Management Board (PMB) Field Replaceable Unit (FRU), 2076139-001-S
Distribution Pattern
Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Argentina, Australia, Bahamas, Bangladesh, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, China, Costa Rica, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, Finland, France, Germany, Guatemala, Hong Kong, India, Indonesia, Iraq, Ireland, Israel, Italy, Japan, Korea (Republic Of), Kuwait, Latvia, Lithuania, Malaysia, Mexico, Moldova, Morocco, Netherlands, New Zealand, Nigeria, North Macedonia, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Taiwan, Thailand, Tunisia, Turkmenistan, Uganda, Ukraine, United Arab Emirates, United Kingdom, Venezuela and Vietnam.
Other Recalls by GE Medical Systems China Co., Ltd.
- Class I: Dangerous 11/14/2025
- Class I: Dangerous 11/14/2025
- Class II: Risk 05/16/2025
- Class II: Risk 05/16/2025
- Class I: Dangerous 03/21/2025