Steris Corporation: Device Recall
Recall #Z-0909-2026 · 11/19/2025
Recall Details
- Recall Number
- Z-0909-2026
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Steris Corporation
- Status
- Ongoing
- Date Initiated
- 11/19/2025
- Location
- Mentor, OH, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 10
Reason for Recall
Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).
Product Description
AMSCO 7052HP Washer/Disinfector Model/Catalog Number: 7052HP Product Description: The AMSCO 7052HP Washer/Disinfector is intended for use in the cleaning and intermediate-level disinfection of soiled reusable simple hard-surfaced rigid surgical instruments (such as forceps and clamps), utensils (such as bedpans, urinals, trays, bowl, basin and kidney dishes), rubber and plastic goods, theatre shoes and other similar and related articles found in healthcare facilities.
Distribution Pattern
Worldwide - US Nationwide distribution in the states of Arizona, Colorado, Florida, Georgia, Illinois, Kentucky, Louisiana, Michigan, New Hampshire, New York, North Carolina, Pennsylvania, Tennessee, Texas, Utah, Victoria, Virginia, Washington, Wisconsin and the countries of Australia.