Illumina, Inc.: Device Recall
Recall #Z-0884-2026 · 10/28/2025
Recall Details
- Recall Number
- Z-0884-2026
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Illumina, Inc.
- Status
- Ongoing
- Date Initiated
- 10/28/2025
- Location
- San Diego, CA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 11916
Reason for Recall
An instrument intended for sequencing DNA libraries with in vitro diagnostic assays is used with reagent kits including flow cells that have flow cell gaskets that have a quality issue, which may lead to flow cell leaks during runs, which could cause run failures, that may result in: No/insufficient data output, instrument fluidics clogs, toxic substance exposure to user.
Product Description
Reagent Kits/ containing flow cells, components of NextSeq 550Dx Instrument: NextSeq 550Dx High Output Reagent Kit v2.5 (75 cycles) IVD, REF: 20028870/NextSeq 550Dx HO FC Cart v2.5, 75 Cycles, REF: 20031098; NextSeq 550Dx High Output Reagent Kit v2.5 (300 cycles) IVD, REF: 20028871/NextSeq 550Dx HO FC Cart v2.5, 300 Cycles, REF: 20026365; CN NextSeq 550Dx High Output Reagent Kit v2.5 (75 cycles) IVD, REF: 20064341/CN NEXTSEQ 550Dx HO FLOW CELL V2.5, REF: 20062316; CN NextSeq 550Dx High Output Reagent Kit v2.5 (300 cycles) IVD, REF: 20064344/CN NEXTSEQ 550Dx HO FLOW CELL V2.5, REF: 20062316; CN NextSeq 550Dx Mid Output Reagent Kit v2.5 (150 cycles) IVD, 20064345/CN NEXTSEQ 550Dx MO FLOW CELL V2.5, REF: 20062530; CN NextSeq 550Dx Mid Output Reagent Kit v2.5 (300 cycles) IVD, 20064346/CN NEXTSEQ 550Dx MO FLOW CELL V2.5, REF: 20062530; /FLOWCELL, FIT FLUIDICS TESTING, REF: 15050205
Distribution Pattern
Worldwide - US Nationwide distribution in the states of PA, NJ, CA, NC, UT, IL, MD, NY, VA, WV, SD, TX, AL, SC, TN, KY, MI and the countries of DK, SA, FR, DE, IT, CN, PL, GB, TH, GR, VN, BE, ES, IL, KR, AU, CH, BG, NL, IE, TW, TR, CZ, HU, NO, FI, JP, SG, LU, AT, AE, ZA, PT, LT, SE.