Abiomed, Inc.: Device Recall
Recall #Z-0829-2025 · 12/12/2024
Recall Details
- Recall Number
- Z-0829-2025
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- Abiomed, Inc.
- Status
- Ongoing
- Date Initiated
- 12/12/2024
- Location
- Danvers, MA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1,560 units
Reason for Recall
Optical Sensors have been damaged due to physical interaction between the inlet and another device resulting in a Placement Signal Not Reliable (PSNR) alarm and loss of Central Venous (CV) Placement Signal, Pulmonary Artery (PA) Placement Signal & Pulmonary Artery Pulse Index (PAPi) metrics. Abiomed is reinforcing that there is a risk of potential interaction between the tip of guidewires, indwelling central venous lines or devices and inlet of the Impella pumps listed above during the insertion, manipulation, and removal of those devices. The interaction may result in optical sensor damage, temporary pump stop, or permanent pump stop.
Product Description
Impella RP with SmartAssist; Product Number: 0046-0035;
Distribution Pattern
Domestic: Nationwide Distribution International: Canada