Boston Scientific Corporation: Device Recall

Recall #Z-0822-2025 · 12/12/2024

Class I: Dangerous

Recall Details

Recall Number
Z-0822-2025
Classification
Class I
Product Type
Device
Recalling Firm
Boston Scientific Corporation
Status
Ongoing
Date Initiated
12/12/2024
Location
Saint Paul, MN, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
156,329 units (59,937 US, 96,792 OUS) in total

Reason for Recall

A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.

Product Description

ACCOLADE DR Pacemaker, SL (Model Number L301), SL MRI (Model Number L311), EL (Model Number L321), and EL MRI (Model Number L331)

Distribution Pattern

US (Nationwide) including Puerto Rico and US Virgin Islands and OUS (foreign) to countries of: Guam and Northern Mariana Islands.

Other Recalls by Boston Scientific Corporation

View all recalls by Boston Scientific Corporation →

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.