Beckman Coulter, Inc.: Device Recall
Recall #Z-0803-2025 · 11/26/2024
Recall Details
- Recall Number
- Z-0803-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Beckman Coulter, Inc.
- Status
- Ongoing
- Date Initiated
- 11/26/2024
- Location
- Chaska, MN, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 362 units
Reason for Recall
Beckman Coulter has identified that some Access PTH reagent packs from lot 339071 may be over- or under-filled in well 0. The dispensed material in well 0 could be above or below specification, risking sealing issues, contamination, instrument errors, delays, or inaccurate results with a negative bias and increase in imprecision. The issue was confirmed on 10 April 2024 by Beckman Coulter through an internal non-conformance (NC-INT-66585). The well 0 scale on the filling line began malfunctioning on the fill immediately before PTH. The malfunction was not noticed until the majority of PTH was filled.
Product Description
Access Intact PTH assay, a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of intact parathyroid hormone (parathyrin, PTH) levels in human serum and plasma using the Access Immunoassay Systems, Catalog Number A16972
Distribution Pattern
Worldwide distribution - US Nationwide and the country of Canada.
Other Recalls by Beckman Coulter, Inc.
- Class II: Risk 12/23/2025
- Class II: Risk 11/26/2025
- Class II: Risk 11/26/2025
- Class II: Risk 11/10/2025
- Class II: Risk 11/07/2025