MEDLINE INDUSTRIES, LP - Northfield: Device Recall

Recall #Z-0764-2025 · 12/16/2024

Class I: Dangerous

Recall Details

Recall Number
Z-0764-2025
Classification
Class I
Product Type
Device
Recalling Firm
MEDLINE INDUSTRIES, LP - Northfield
Status
Ongoing
Date Initiated
12/16/2024
Location
Northfield, IL, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13,600 units

Reason for Recall

Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.

Product Description

Medline Fluid Delivery Set with Drip Chamber: 1) FDS,135,CHMN,W/F,MLL,RC(2),PG, Item Number 99000069; 2) FDS,135,NV,MICRO-CHAMBER,W/FILTER,BLUE, Item Number 949000222; 3) FDS,135,NV,MICRO-CHAMBER,W/FILTER,GREEN, Item Number 949000224; 4) FDS,135,NV,MICRO-CHAMBER,W/FILTER,RED, Item Number 949000221; 5) FDS,135,NV,MICRO-CHAMBER,W/FILTER,YELLOW, Item Number 949000223

Distribution Pattern

US Nationwide Distribution

Other Recalls by MEDLINE INDUSTRIES, LP - Northfield

View all recalls by MEDLINE INDUSTRIES, LP - Northfield →

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.