MEDLINE INDUSTRIES, LP - Northfield: Device Recall
Recall #Z-0764-2025 · 12/16/2024
Class I: Dangerous
Recall Details
- Recall Number
- Z-0764-2025
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- MEDLINE INDUSTRIES, LP - Northfield
- Status
- Ongoing
- Date Initiated
- 12/16/2024
- Location
- Northfield, IL, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 13,600 units
Reason for Recall
Fluid Delivery Sets were incorrectly assembled with a white macro drip chamber instead of the required grey micro drip chamber. The macro drip chamber delivers three times more fluid per drop than the micro drip chamber.
Product Description
Medline Fluid Delivery Set with Drip Chamber: 1) FDS,135,CHMN,W/F,MLL,RC(2),PG, Item Number 99000069; 2) FDS,135,NV,MICRO-CHAMBER,W/FILTER,BLUE, Item Number 949000222; 3) FDS,135,NV,MICRO-CHAMBER,W/FILTER,GREEN, Item Number 949000224; 4) FDS,135,NV,MICRO-CHAMBER,W/FILTER,RED, Item Number 949000221; 5) FDS,135,NV,MICRO-CHAMBER,W/FILTER,YELLOW, Item Number 949000223
Distribution Pattern
US Nationwide Distribution