Beckman Coulter, Inc.: Device Recall

Recall #Z-0727-2025 · 11/18/2024

Class II: Risk

Recall Details

Recall Number
Z-0727-2025
Classification
Class II
Product Type
Device
Recalling Firm
Beckman Coulter, Inc.
Status
Ongoing
Date Initiated
11/18/2024
Location
Chaska, MN, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,109,804 units

Reason for Recall

Beckman Coulter, Inc. is recalling their Access hsTnI assay, a medical device by correction. The reason for the recall is: high troponin antigen (>55,000 pg/mL) can contaminate an analyzers probe and cause inter-assay carryover due to insufficient washing. The contamination can cause a falsely elevated Access hsTnI result in subsequent sample(s) which can impact patient care (potentially but not limited to unnecessary coronary imaging or diagnostic catheterization) if the result is near the medical decision points.

Product Description

Access hsTnI Reagent, REF B52699, Immunoassay Method, Troponin Subunit. For the quantitative determination of cardiac troponin I (cTnI) levels in human serum and plasma.

Distribution Pattern

Worldwide