Philips North America Llc: Device Recall
Recall #Z-0686-2025 · 11/20/2024
Recall Details
- Recall Number
- Z-0686-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Philips North America Llc
- Status
- Ongoing
- Date Initiated
- 11/20/2024
- Location
- Cambridge, MA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4,811 units
Reason for Recall
The following four issues are included: 1. Certain Channel Settings may result in an Access Point preventing a connected client device from roaming to another Access Point potentially causing the client device to appear to drop out from the network. 2. if the Access Point experiences a disconnection from the Access Point Controller exceeding 4 minutes, the Access Point may not transition through the desired state to all for clean reconnection of the client radio link. 3. Potential to experience higher than expected data dropouts, device disassociations/reassociations for their wireless clients, and frequent "Loss of AP" alerts. 4. Potential for the Smart-hopping Access Point is affected by a software issue where the Access Point will resent after 82.85 days when the frame counter reaches its limit. This causes the latest synchronization check to fail and potential for loss of patient data.
Product Description
Philips Smart-Hopping 2.0 AP 1.4 GHz. Model Number: 867216
Distribution Pattern
Nationwide distribution