Philips North America Llc: Device Recall

Recall #Z-0686-2025 · 11/20/2024

Class II: Risk

Recall Details

Recall Number
Z-0686-2025
Classification
Class II
Product Type
Device
Recalling Firm
Philips North America Llc
Status
Ongoing
Date Initiated
11/20/2024
Location
Cambridge, MA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
4,811 units

Reason for Recall

The following four issues are included: 1. Certain Channel Settings may result in an Access Point preventing a connected client device from roaming to another Access Point potentially causing the client device to appear to drop out from the network. 2. if the Access Point experiences a disconnection from the Access Point Controller exceeding 4 minutes, the Access Point may not transition through the desired state to all for clean reconnection of the client radio link. 3. Potential to experience higher than expected data dropouts, device disassociations/reassociations for their wireless clients, and frequent "Loss of AP" alerts. 4. Potential for the Smart-hopping Access Point is affected by a software issue where the Access Point will resent after 82.85 days when the frame counter reaches its limit. This causes the latest synchronization check to fail and potential for loss of patient data.

Product Description

Philips Smart-Hopping 2.0 AP 1.4 GHz. Model Number: 867216

Distribution Pattern

Nationwide distribution