Spacelabs Healthcare, Ltd.: Device Recall

Recall #Z-0685-2025 · 11/25/2024

Class II: Risk

Recall Details

Recall Number
Z-0685-2025
Classification
Class II
Product Type
Device
Recalling Firm
Spacelabs Healthcare, Ltd.
Status
Ongoing
Date Initiated
11/25/2024
Location
Hertford, United Kingdom
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,779 devices

Reason for Recall

Due to a manufacturing error the incorrect device serial number was potentially programmed into the holter recorder. The physical serial number label on the device might not match the serial number programmed into the device.

Product Description

Eclipse Mini REF 98900 The Eclipse MINI Model 98900 is a portable non-invasive continuous ambulatory ECG patch recorder intended to record the patient s electrocardiogram.

Distribution Pattern

U.S.: AR, CA, CO, DE, FL, FA, LA, MA, MI, MN, MT, NC, NE, NU OH, OR, PA, Puerto Rico, TN, TX, UT, VA, WA, and WI. O.U.S.: Argentina Denmark Kuwait Serbia Australia Dominican Republic Lebanon Singapore Austria Ecuador Lithuania Spain Belgium El Salvador Martinique Sweden Benin France Mexico Switzerland Bolivia Germany Netherlands Taiwan Brazil Greece Oman Trinidad and Tobago Brunei Darussalam Guatemala Panama Turkey Canada Honduras Peru United Arab Emirates Chile Hong Kong Poland United Kingdom China Iceland Portugal Columbia India Venezuela Congo Ireland Qatar Viet Nam Costa Rica Israel Republic of Macedonia Croatia Italy Roumanie Czech Republic Korea, Republic of Saudi Arabia