Nova Biomedical Corporation: Device Recall
Recall #Z-0681-2025 · 11/20/2024
Recall Details
- Recall Number
- Z-0681-2025
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- Nova Biomedical Corporation
- Status
- Ongoing
- Date Initiated
- 11/20/2024
- Location
- Waltham, MA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2,365 units
Reason for Recall
A software bug within the Gen 2 StatStrip Hospital Meter firmware leads to transmission of erroneous glucose patient test results to a healthcare institution's data management system (DMS) at healthcare sites using WiFi if a user navigates to the Review Results screen while the meter is in the process of transmitting the current glucose test result to the DMS. The risk of this software defect is erroneously high or low glucose results being documented in a patient's medical record, which may lead to incorrect treatment.
Product Description
StatStrip Glucose Hospital Meter System. Model/Catalog Number: 63685. Prescription use blood glucose meter for near-patient testing.
Distribution Pattern
Distribution US Nationwide including Puerto Rico, Germany, Italy, Israel, Switzerland, Australia, Great Britain, and the Netherlands.
Other Recalls by Nova Biomedical Corporation
- Class I: Dangerous 11/20/2024
- Class I: Dangerous 11/20/2024