B Braun Medical Inc: Device Recall
Recall #Z-0669-2026 · 10/29/2025
Recall Details
- Recall Number
- Z-0669-2026
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- B Braun Medical Inc
- Status
- Ongoing
- Date Initiated
- 10/29/2025
- Location
- Bethlehem, PA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 190,625 units
Reason for Recall
Potential for backflow of medication from secondary (piggyback) IV containers into primary IV containers and the inability to prime (occlusion).
Product Description
IV Administration Set utilized in gravity IV administration sets and pump administration sets utilized with BBMI's Infusomat Space Large Volume Pump, Outlook Pump, and Vista Basic Pump. Labeled as the following with corresponding catalog number: 1. IV ADMIN SET 15 DROP W/ CARESITE EXT SET; Catalog Number: 490528. 2. 111 IN. IV ADMIN SET W/ 12 IN. Y-EXT SET; Catalog Number: 490550. 3. IV ADMIN SET, 125 IN. w/ EXT SET 9 IN.; Catalog Number: 490605. 4. IV SET 15 DR 3CARESITE 2 ULTRAPORT; Catalog Number: 490616. 5. IV ADMIN SET, 2 CARSITE LADS; Catalog Number: 490654.
Distribution Pattern
Worldwide distribution - US distribution and international distribution to Canada, Germany, Guatemala, and Singapore.