Shimadzu Corp. Analytical & Measuring Instruments Div.: Device Recall
Recall #Z-0554-2025 · 10/09/2024
Class II: Risk
Recall Details
- Recall Number
- Z-0554-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Shimadzu Corp. Analytical & Measuring Instruments Div.
- Status
- Ongoing
- Date Initiated
- 10/09/2024
- Location
- Kyoto, Japan
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 48
Reason for Recall
There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.
Product Description
Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL
Distribution Pattern
Worldwide - US Nationwide distribution in the states of DE, PA and the country of DE.