Shimadzu Corp. Analytical & Measuring Instruments Div.: Device Recall

Recall #Z-0554-2025 · 10/09/2024

Class II: Risk

Recall Details

Recall Number
Z-0554-2025
Classification
Class II
Product Type
Device
Recalling Firm
Shimadzu Corp. Analytical & Measuring Instruments Div.
Status
Ongoing
Date Initiated
10/09/2024
Location
Kyoto, Japan
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
48

Reason for Recall

There have been cases of failure of the main board due to condensation water occurring in the gap between thermal insulation material on the underside of the main base dropping onto the main board The symptoms and errors differ depending on the position where the water drops adhere. To date there have been examples of reports of inter-board communication errors, HOME POSITION error or SLIP error of XYZ drive units. The failure could result in delay of patient sample testing.

Product Description

Autosampler, Models: SIL-40C XR CL, SIL-40C X3 CL

Distribution Pattern

Worldwide - US Nationwide distribution in the states of DE, PA and the country of DE.