Shimadzu Corp. Analytical & Measuring Instruments Div.: Device Recall
Recall #Z-0553-2025 · 10/09/2024
Recall Details
- Recall Number
- Z-0553-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Shimadzu Corp. Analytical & Measuring Instruments Div.
- Status
- Ongoing
- Date Initiated
- 10/09/2024
- Location
- Kyoto, Japan
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 179
Reason for Recall
There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and adjacent cables are partly charred. In the minor case, the Y motor and its surrounding die cast part become rusty. These failures could result in delay in patient sample testing.
Product Description
Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL
Distribution Pattern
Worldwide - US Nationwide distribution in the states of MS, AR, TX, CA, MA, MI, FL, OK, WI, IN, MN, OH, NJ, NY, IL, KY, NV, HI, SD, ID, MI, UT, WA, MO, VA, GA and the countries of DE, CN, JP, UY.
Other Recalls by Shimadzu Corp. Analytical & Measuring Instruments Div.
- Class II: Risk 10/09/2024
View all recalls by Shimadzu Corp. Analytical & Measuring Instruments Div. →