Shimadzu Corp. Analytical & Measuring Instruments Div.: Device Recall

Recall #Z-0553-2025 · 10/09/2024

Class II: Risk

Recall Details

Recall Number
Z-0553-2025
Classification
Class II
Product Type
Device
Recalling Firm
Shimadzu Corp. Analytical & Measuring Instruments Div.
Status
Ongoing
Date Initiated
10/09/2024
Location
Kyoto, Japan
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
179

Reason for Recall

There have been cases of failure of the main board due to condensation water drained from the Peltier dehumidifier, leaked out from the tray, resulting in a short circuit. In the serious case, the main board and adjacent cables are partly charred. In the minor case, the Y motor and its surrounding die cast part become rusty. These failures could result in delay in patient sample testing.

Product Description

Autosampler, Models: SIL-20ACXR CL, SIL-20ACHT CL, SIL-30AC CL

Distribution Pattern

Worldwide - US Nationwide distribution in the states of MS, AR, TX, CA, MA, MI, FL, OK, WI, IN, MN, OH, NJ, NY, IL, KY, NV, HI, SD, ID, MI, UT, WA, MO, VA, GA and the countries of DE, CN, JP, UY.