Abbott Medical: Device Recall

Recall #Z-0552-2025 · 10/01/2024

Class III: Low Risk

Recall Details

Recall Number
Z-0552-2025
Classification
Class III
Product Type
Device
Recalling Firm
Abbott Medical
Status
Ongoing
Date Initiated
10/01/2024
Location
Sylmar, CA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
Pending

Reason for Recall

Due to heart transmitter not being able to upgrade to the current software version in the field after experiencing a power interruption during a prior over-the-wire software download.

Product Description

Merlin@home with the MerlinOnDemand capability Transmitter REF EX1150 Merlin@home with the MerlinOnDemand" (MoD) capability is a transmitter configured for multiple patient use. It can receive information from an implantable heart device in unpaired mode. With MoD capability, the transmitter will retrieve full follow-up data or retrieve data for display on the Merlin.net" Patient Care Network (PCN). The MoD capability enables clinicians to do the following: . support remote care patients in emergency rooms and heart failure clinics . triage patients for in clinic follow-ups . support patients in a clinical setting with Merlin.net PCN

Distribution Pattern

Worldwide - U.S. Nationwide distribution in the states of ALABAMA, ALASKA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, ILLINOIS, INDIANA, KENTUCKY, MICHIGAN, MISSISSIPPI, MISSOURI, MONTANA, NEBRASKA, NEW JERSEY, NEW YORK, NORTH CAROLINA, OHIO, OREGON, PENNSYLVANIA, SOUTH CAROLINA, TENNESSEE, TEXAS, VIRGINIA, WASHINGTON and WISCONSIN. The country of Australia.

Other Recalls by Abbott Medical