Medline Industries, LP: Device Recall

Recall #Z-0524-2026 · 09/30/2025

Class II: Risk

Recall Details

Recall Number
Z-0524-2026
Classification
Class II
Product Type
Device
Recalling Firm
Medline Industries, LP
Status
Ongoing
Date Initiated
09/30/2025
Location
Northfield, IL, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2844 kits

Reason for Recall

Medline Industries, LP is issuing a recall for specific item(s) and lot(s) of Medline Kits containing SafeAir Smoke Evacuation Pencils. Stryker Instruments has expanded the scope of their recall of SafeAir Smoke Evacuation Pencils previously initiated in October 2024, adding an additional 47 lots to the scope of the recall. The initial recall was due to the potential for product to activate without manual input when plugged into a power source or remain active after buttons are released.

Product Description

MEDLINE convenience kits labeled as: 1) GYN LAPAROSCOPY CDS #34-RF, REF CDS920099S; 2) GYN LAPAROSCOPY CDS-LF, REF CDS983764J; 3) ROBOTIC, REF CDS984262N; 4) ROBOTIC, REF CDS984262P; 5) ROBOTIC UROLOGY-RF, REF DYNJ38844N; 6) DA VINCI PROSTATE/COLPOPEXY, REF DYNJ44864S; 7) LAP-GYN PACK, REF DYNJ44876M; 8) ROBOT PACK, REF DYNJ45184F; 9) GYN LAP PACK, REF DYNJ52524D; 10) GYN LAP PACK, REF DYNJ64234C; 11) GYN DAVINCI PACK, REF DYNJ65373A; 12) ROBOT PACK, REF DYNJ67190C; 13) ROBOTIC PACK, REF DYNJ86154; 14) GYN LAPAROTOMY, REF DYNJ900296K; 15) ROBOT THOR, REF DYNJ900300K; 16) ROBOT URO, REF DYNJ900301K; 17) SM DAVINCI, REF DYNJ903539G; 18) ROBOTIC PACK-LF, REF DYNJ904250I; 19) ROBOTIC-LF, REF DYNJ907910B; 20) GP-ROBOTIC PACK-LF, REF PHS504647004G.

Distribution Pattern

US Nationwide distribution.