Siemens Healthcare Diagnostics, Inc.: Device Recall

Recall #Z-0492-2026 · 10/06/2025

Class II: Risk

Recall Details

Recall Number
Z-0492-2026
Classification
Class II
Product Type
Device
Recalling Firm
Siemens Healthcare Diagnostics, Inc.
Status
Ongoing
Date Initiated
10/06/2025
Location
East Walpole, MA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2802 units (OUS only)

Reason for Recall

The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.

Product Description

3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1), 3gAllergy Specific IgE Control Antibody (L2UNS2); Siemens Material Number (SMN): 10380875; Catalog Number: L2KUN6;

Distribution Pattern

Worldwide - US Nationwide distribution in the state of CA and the countries of Argentina, Australia, Austria, Belgium, Bosnia Herzeg. , Brazil, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Estonia, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Israel, Italy, Latvia, Macedonia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland, Portugal, Romania, Russian Fed., Serbia, Slovakia, Spain, Sweden, Turkey, U.A.E., Ukraine, United Kingdom, Uruguay.