Siemens Healthcare Diagnostics, Inc.: Device Recall
Recall #Z-0492-2026 · 10/06/2025
Recall Details
- Recall Number
- Z-0492-2026
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Siemens Healthcare Diagnostics, Inc.
- Status
- Ongoing
- Date Initiated
- 10/06/2025
- Location
- East Walpole, MA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2802 units (OUS only)
Reason for Recall
The barcode orientation on affected devices causes incorrect scanning order of tubes within the allergen holder wedge. When using an IMMULITE 2000 / IMMULITE 2000 XPi instrument in combination with a Zebra barcode scanner, the instrument will not correctly recognize the position of tubes within the allergen holder wedge.
Product Description
3gAllergy Specific IgE Universal Kit; Kit Component: 3gAllergy Specific IgE Adjustor Antibody (L2UNS1), 3gAllergy Specific IgE Control Antibody (L2UNS2); Siemens Material Number (SMN): 10380875; Catalog Number: L2KUN6;
Distribution Pattern
Worldwide - US Nationwide distribution in the state of CA and the countries of Argentina, Australia, Austria, Belgium, Bosnia Herzeg. , Brazil, Canary Islands, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Estonia, France, Germany, Greece, Guatemala, Hong Kong, Hungary, Israel, Italy, Latvia, Macedonia, Mexico, Netherlands, Norway, Pakistan, Panama, Poland, Portugal, Romania, Russian Fed., Serbia, Slovakia, Spain, Sweden, Turkey, U.A.E., Ukraine, United Kingdom, Uruguay.