Smiths Medical ASD, Inc.: Device Recall

Recall #Z-0466-2025 · 06/26/2024

Class II: Risk

Recall Details

Recall Number
Z-0466-2025
Classification
Class II
Product Type
Device
Recalling Firm
Smiths Medical ASD, Inc.
Status
Ongoing
Date Initiated
06/26/2024
Location
Minneapolis, MN, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
156,400 AC adapters

Reason for Recall

The adapter input plug can become damaged or broken. If the input plug is damaged, the metal contacts to the body of the AC adapter may be exposed or one or more of the AC mains prongs may separate from the input plug.

Product Description

smiths medical CADD-Solis Li-ion Rechargeable Battery Packs AC adapter, Model Number REF 21-0270-25, for use with the CADD-Solis infusion pumps

Distribution Pattern

Worldwide distribution.

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.