GE Healthcare Finland Oy: Device Recall

Recall #Z-0457-2026 · 09/26/2025

Class II: Risk

Recall Details

Recall Number
Z-0457-2026
Classification
Class II
Product Type
Device
Recalling Firm
GE Healthcare Finland Oy
Status
Ongoing
Date Initiated
09/26/2025
Location
Helsinki, Finland
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2857

Reason for Recall

Potential loss of monitoring on affected monitors when powered by certain batteries. Affected batteries can lose capability to hold charge and provide back up power. If this situation occurs, it will cause loss of patient monitoring and potentially delay recognizing patient status changes requiring treatment.

Product Description

BATTERY FLEX-3S2P 10.8V 18650 LI-ION SMBUS used with affected patient monitors Model No 2062895-001 Manufacturer Part Number U80296-1R01, revision S

Distribution Pattern

Worldwide - US Nationwide distribution.