Olympus Corporation of the Americas: Device Recall

Recall #Z-0452-2026 · 10/08/2025

Class II: Risk

Recall Details

Recall Number
Z-0452-2026
Classification
Class II
Product Type
Device
Recalling Firm
Olympus Corporation of the Americas
Status
Ongoing
Date Initiated
10/08/2025
Location
Center Valley, PA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1528 units

Reason for Recall

The device may either fail to start up, or the transducer may start briefly and then stop, accompanied by an error light on the generator. In addition, the body of the transducer handpiece may gradually increase in temperature during clinical use.

Product Description

Brand Name: Lithotriptor, Ultrasonic Product Name: ShockPulse Lithotripsy Transducer Model/Catalog Number: SPL-T Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones. Component: ShockPulse Lithotripsy System

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Canada, India, Taiwan, Germany, Australia, Singapore.

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.