Medline Industries, LP: Device Recall

Recall #Z-0448-2026 · 10/10/2025

Class I: Dangerous

Recall Details

Recall Number
Z-0448-2026
Classification
Class I
Product Type
Device
Recalling Firm
Medline Industries, LP
Status
Ongoing
Date Initiated
10/10/2025
Location
Northfield, IL, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
96,406 total units

Reason for Recall

Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.

Product Description

Convenience Kits: DYNJ901048N ORTHO SPINE ST CLAIR DYNJ909123F UROLOGY MINOR DYNJ909132F SET UP DYNJ909308D HIP PINNING

Distribution Pattern

International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;