Medline Industries, LP: Device Recall
Recall #Z-0444-2026 · 10/10/2025
Class I: Dangerous
Recall Details
- Recall Number
- Z-0444-2026
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- Medline Industries, LP
- Status
- Ongoing
- Date Initiated
- 10/10/2025
- Location
- Northfield, IL, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 96,406 total units
Reason for Recall
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Product Description
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; OPEN HEART SUPPLEMENT DYNJ909116D; VASCULAR DYNJ909124F
Distribution Pattern
International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;