Beckman Coulter, Inc.: Device Recall
Recall #Z-0412-2026 · 09/08/2025
Recall Details
- Recall Number
- Z-0412-2026
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Beckman Coulter, Inc.
- Status
- Ongoing
- Date Initiated
- 09/08/2025
- Location
- Miami, FL, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 108 units
Reason for Recall
Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters, and E/1300/2093 microliters. Device software does not rock Cassette E, used with 15mmX92mm Sarstedt tubes. May lead to incorrect diagnosis or error in patient management/treatment.
Product Description
CellMek SPS Sample Preparation System, REF: C44603. Software versions 2.2.122.0 and prior, and Instructions for Use (IFU) revisions AD and prior. These CellMek devices include Cassette E.
Distribution Pattern
Worldwide - US Nationwide distribution in the states of AL, AR, FL, IL, IN, MA, ME, MI, MN, MO, NY, OR, TN, UT, WA and the countries Australia, Austria, Belgium, Brazil, Canada, China, France, Germany, Greece, Hong Kong, Israel, Italy, Japan, Korea, Republic of, Mexico, Morocco, Netherlands, Poland, Portugal, Puerto Rico, South Africa, Spain, Switzerland, United Kingdom of Great Britain and Northern Ireland.