Beckman Coulter, Inc.: Device Recall

Recall #Z-0412-2026 · 09/08/2025

Class II: Risk

Recall Details

Recall Number
Z-0412-2026
Classification
Class II
Product Type
Device
Recalling Firm
Beckman Coulter, Inc.
Status
Ongoing
Date Initiated
09/08/2025
Location
Miami, FL, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
108 units

Reason for Recall

Automated pipetting, diluting and specimen processing workstations for flow cytometric analysis device instructions for use list incorrect dead volumes: Cassette Type/List volume/correct volume, B/750/1582 microliters, and E/1300/2093 microliters. Device software does not rock Cassette E, used with 15mmX92mm Sarstedt tubes. May lead to incorrect diagnosis or error in patient management/treatment.

Product Description

CellMek SPS Sample Preparation System, REF: C44603. Software versions 2.2.122.0 and prior, and Instructions for Use (IFU) revisions AD and prior. These CellMek devices include Cassette E.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of AL, AR, FL, IL, IN, MA, ME, MI, MN, MO, NY, OR, TN, UT, WA and the countries Australia, Austria, Belgium, Brazil, Canada, China, France, Germany, Greece, Hong Kong, Israel, Italy, Japan, Korea, Republic of, Mexico, Morocco, Netherlands, Poland, Portugal, Puerto Rico, South Africa, Spain, Switzerland, United Kingdom of Great Britain and Northern Ireland.