Stryker Corporation: Device Recall
Recall #Z-0368-2025 · 10/04/2024
Recall Details
- Recall Number
- Z-0368-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Stryker Corporation
- Status
- Ongoing
- Date Initiated
- 10/04/2024
- Location
- Portage, MI, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1,996,010 eaches
Reason for Recall
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
Product Description
Stryker Neptune SafeAir Smoke Evacuation Pencil, Coated, 70mm Blade, Push Button Switch - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-046-000
Distribution Pattern
Worldwide distribution - US Nationwide and the countries of Canada, Japan, Australia.