Baxter Healthcare Corporation: Device Recall

Recall #Z-0324-2025 · 10/11/2024

Class II: Risk

Recall Details

Recall Number
Z-0324-2025
Classification
Class II
Product Type
Device
Recalling Firm
Baxter Healthcare Corporation
Status
Ongoing
Date Initiated
10/11/2024
Location
Deerfield, IL, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1369 units

Reason for Recall

There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.

Product Description

Hillrom Welch Allyn Diagnostic Cardiology Suite (formerly Connex Cardio); Diagnostic spirometer

Distribution Pattern

Worldwide distribution.