Baxter Healthcare Corporation: Device Recall
Recall #Z-0322-2025 · 10/11/2024
Class II: Risk
Recall Details
- Recall Number
- Z-0322-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Baxter Healthcare Corporation
- Status
- Ongoing
- Date Initiated
- 10/11/2024
- Location
- Deerfield, IL, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 399 units
Reason for Recall
There is the potential for exam files being assigned duplicate Unique Identifiers (UIDs),. If the system receiving the DICOM file (e.g., Picture Archiving and Communication System (PACS)) relies solely on the UID to accept exams, this could lead to a mismatch of the patients identification with their physiological data.
Product Description
Hillrom Welch Allyn RScribe 12-Lead Electrocardiograph System; Electrocardiograph
Distribution Pattern
Worldwide distribution.