Beckman Coulter, Inc.: Device Recall

Recall #Z-0313-2025 · 09/30/2024

Class II: Risk

Recall Details

Recall Number
Z-0313-2025
Classification
Class II
Product Type
Device
Recalling Firm
Beckman Coulter, Inc.
Status
Ongoing
Date Initiated
09/30/2024
Location
Chaska, MN, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
149 units

Reason for Recall

The Access Cortisol assay protocol file (APF) applies an incorrect calibration curve acceptance parameter at the S1 calibrator level that can lead to a high rate of calibration failures which can vary by reagent lot number. Calibration failure would lead to delay of measurement results reporting and potential injuries.

Product Description

The Access Cortisol assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative determination of cortisol levels in human serum, plasma (heparin, EDTA) and urine using the Access Immunoassay Systems.

Distribution Pattern

Worldwide distribution.