Baxter Healthcare Corporation: Device Recall

Recall #Z-0228-2025 · 09/18/2024

Class II: Risk

Recall Details

Recall Number
Z-0228-2025
Classification
Class II
Product Type
Device
Recalling Firm
Baxter Healthcare Corporation
Status
Ongoing
Date Initiated
09/18/2024
Location
Deerfield, IL, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,967 beds

Reason for Recall

The beds have a potential for an interface issue with the Rauland Responder 5 Nurse Call System where the bed exit system will alert at the bed, but may fail to send a remote alert through the nurse call system.

Product Description

a. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, Product codes: P7900B000001, P7900B100010, P7900B100024, P7900B100208, P7900B100273, P7900B100304, P7900B000005, P7900B100011, P7900B100025, P7900B100212, P7900B100277, P7900B100307, P7900B000010, P7900B100017, P7900B100026, P7900B100222, P7900B100281, P7900B1SPL05, P7900B000011, P7900B100018, P7900B100028, P7900B100223, P7900B100287, P7900B1SPL06, P7900B000020, P7900B100019, P7900B100039, P7900B100224, P7900B100288, P7900B000022, P7900B100020, P7900B100090, P7900B100231, P7900B100292, P7900B100001, P7900B100021, P7900B100161, P7900B100237, P7900B100293, P7900B100005, P7900B100022, P7900B100203, P7900B100242, P7900B100294, P7900B100009, P7900B100023, P7900B100205, P7900B100264, and P7900B100296. b. Hillrom Centrella Smart+ Medical-Surgical Hospital Bed, Product codes: P7900A000001, P7900A000010, P7900A000019, P7900A000039, P7900A000005, P7900A000011, P7900A000022, and P7900A000063.

Distribution Pattern

Worldwide distribution - US Nationwide. There was government and military distribution. The countries of Canada and Australia.