Olympus Corporation of the Americas: Device Recall

Recall #Z-0225-2025 · 09/13/2024

Class II: Risk

Recall Details

Recall Number
Z-0225-2025
Classification
Class II
Product Type
Device
Recalling Firm
Olympus Corporation of the Americas
Status
Ongoing
Date Initiated
09/13/2024
Location
Center Valley, PA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
209 units

Reason for Recall

Mislabeled Glo Cath Label (attached to the device) states: 11 mm, 12 mm, 13 mm however the balloon is 8.5mm, 9.5mm and 10.5 mm and may result in prolonged surgery. All other labeling is correct for the balloon, including the Shelf Box with Front Box Label and Circular Star Label and the balloon Pouch

Product Description

EZDilate Wire Guided Balloon- Indicated for use in adult and adolescent (> 12 years) populations to endoscopically dilate strictures of the alimentary tract. It is also indicated in adults for endoscopic dilatation of the Sphincter of Oddi with or without prior sphincterotomy. Model/Catalog Number: BD-410X-1055

Distribution Pattern

Worldwide Distribution: US distribution to states of: CA, FL, IL, NJ, NY, PA, TN, VA, WI; and Foreign OUS to countries of: CA, DE, IN, SG, Latin America.