Polymer Technology Systems, Inc. dba PTS Diagnostics, Inc.: Device Recall

Recall #Z-0223-2025 · 09/18/2024

Class II: Risk

Recall Details

Recall Number
Z-0223-2025
Classification
Class II
Product Type
Device
Recalling Firm
Polymer Technology Systems, Inc. dba PTS Diagnostics, Inc.
Status
Completed
Date Initiated
09/18/2024
Location
Whitestown, IN, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,452 units

Reason for Recall

The reason for this recall is PTS Diagnostics, Inc. received complaints of non-matching cartridge and analyzer codes within the product test system box. Use of cartridges and analyzers with non-matching codes may result in inaccurate test results as components within each lot number are meant to be used together, and not mixed between lots. Therefore, the test results may possibly lead to Hemoglobin A1C positive or negative bias results. While the product is for monitoring only, over recovery with a higher than accurate value can be an indication for increased therapy; such as physical activity, weight reduction, meal management, pharmaceutical therapy.

Product Description

A1CNow+ Multi-Test HbA1c System consists of (1) a semi-disposable plastic-encased device (the monitor), (2) a plastic cartridge enclosing dry reagent strips, and (3) a sample dilution kit for: collecting the blood sample, mixing the sample with the required pre-treatment solution, and delivering the sample to the cartridge.

Distribution Pattern

Worldwide distribution - US Nationwide and the countries of Japan, South Africa.