Beckman Coulter, Inc.: Device Recall
Recall #Z-0216-2026 · 09/19/2025
Class II: Risk
Recall Details
- Recall Number
- Z-0216-2026
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Beckman Coulter, Inc.
- Status
- Ongoing
- Date Initiated
- 09/19/2025
- Location
- Chaska, MN, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 571 units
Reason for Recall
Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.
Product Description
DxI 9000 Access Immunoassay Analyzer, Part Number C11137
Distribution Pattern
Worldwide - US Nationwide distribution.