Beckman Coulter, Inc.: Device Recall

Recall #Z-0215-2026 · 09/19/2025

Class II: Risk

Recall Details

Recall Number
Z-0215-2026
Classification
Class II
Product Type
Device
Recalling Firm
Beckman Coulter, Inc.
Status
Ongoing
Date Initiated
09/19/2025
Location
Chaska, MN, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
160 units

Reason for Recall

Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.

Product Description

DxC 500i Clinical Analyzer Modules, DxC 500 AU Module w/ISE, DxC 500i (Part Number C63522), DxC 500 AU Module, DxC 500i (Part Number C63521), and Access 2 Module, DxC 500i (Part Number C13252)

Distribution Pattern

Worldwide - US Nationwide distribution.