Wipro GE Healthcare Private Ltd.: Device Recall

Recall #Z-0214-2026 · 09/05/2025

Class II: Risk

Recall Details

Recall Number
Z-0214-2026
Classification
Class II
Product Type
Device
Recalling Firm
Wipro GE Healthcare Private Ltd.
Status
Ongoing
Date Initiated
09/05/2025
Location
Bengaluru, India
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2306 units

Reason for Recall

GE HealthCare has identified an issue affecting Carescape" Central Station (CSCS) with software version V3.0.5, which can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.

Product Description

GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5, Part Number 5867474-03; Medical Display, cathode ray tube

Distribution Pattern

Worldwide distribution.