Wipro GE Healthcare Private Ltd.: Device Recall
Recall #Z-0213-2026 · 09/05/2025
Class II: Risk
Recall Details
- Recall Number
- Z-0213-2026
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Wipro GE Healthcare Private Ltd.
- Status
- Ongoing
- Date Initiated
- 09/05/2025
- Location
- Bengaluru, India
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 173 units
Reason for Recall
GE HealthCare has identified an issue affecting Carescape" Central Station (CSCS) with software version V3.0.5, which can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.
Product Description
GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5; Part Number 5867474-04; Medical Display, cathode ray tube
Distribution Pattern
Worldwide distribution.