Smiths Medical ASD, Inc.: Device Recall

Recall #Z-0204-2025 · 09/24/2024

Class II: Risk

Recall Details

Recall Number
Z-0204-2025
Classification
Class II
Product Type
Device
Recalling Firm
Smiths Medical ASD, Inc.
Status
Ongoing
Date Initiated
09/24/2024
Location
Minneapolis, MN, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1640 units

Reason for Recall

Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.

Product Description

medex TranStar MONITORING SET W/SAMPLING PORT 10/EA, Product Code REF MXG300260

Distribution Pattern

Worldwide distribution.

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.