Olympus Corporation of the Americas: Device Recall
Recall #Z-0181-2025 · 08/21/2024
Class II: Risk
Recall Details
- Recall Number
- Z-0181-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- Olympus Corporation of the Americas
- Status
- Ongoing
- Date Initiated
- 08/21/2024
- Location
- Center Valley, PA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1407 units
Reason for Recall
It was discovered during device performance testing that the CCD imaging sensors were programmed with the incorrect color correction data and therefore, specifications are not met.
Product Description
Olympus Gastrointestinal Videoscope, Model/Catalog Number: GIF-1100
Distribution Pattern
US Nationwide distribution.