American Contract Systems Inc.: Device Recall

Recall #Z-0172-2026 · 09/04/2025

Class II: Risk

Recall Details

Recall Number
Z-0172-2026
Classification
Class II
Product Type
Device
Recalling Firm
American Contract Systems Inc.
Status
Ongoing
Date Initiated
09/04/2025
Location
Grimes, IA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
32,433 total

Reason for Recall

Identified products were re-gassed (sterilized more than once) after a nonconformance occurred during the initial Ethylene Oxide (EO) gas injection process. These products have not been validated for exposure to multiple sterilization cycles. Therefore, product quality, safety, and effectiveness cannot be confirmed for product that was re-gassed.

Product Description

Medical convenience kits Item number/ Product description AKGN82E GENERAL AMCB08Y CHEST BREAST PACK ANBP10AC BASIC PACK - 205939 ANIR77AM INTERVENTIONAL RADIOLOGY PACK - 205953 CFLA39G LAPAROTOMY PACK CIBL33I BASIC LAP PACK EVMN24E MINOR ESC EVST48E STARTER FDTA11AA T & A PACK FHHY25P HYSTEROSCOPY PACK HGML99G MAJOR LAP PACK HGXR41H X-RAY PACK HISU54M SET-UP PACK IHAA19AP PK A A A IHMP89K MINOR PROCEDURE PACK IHPD14AM PK PEDIATRIC LKVG10O VAG DELIVERY LMAG20G VASCULAR PACK LMLE46AA EXTREMITY LOWER PK MIEN58A DTC GENERAL ENDOSCOPY PACK(PS 209468) NUMB04H MAJOR BASIN SET 314666 OWBS11U BASIN SET - 206021 RCMS71B MIS SPECIALTY RCH SFMB55X MINOR BASIN - 208293 STBA09F BASIN SET TNBS21O BASIN SET 612833 UDBH37AX BASIC HEART PACK - 206044 UDMA50X MINOR ADULT - 206067 UICR54W ACS CRS PACK (PS055042) UIPC89D PROCUREMENT PACK UIPT13F PITCHER (PS 907953) UISB96AB SMALL BACK TABLE SETUP (PS 907419)

Distribution Pattern

US Nationwide distribution in the states of SD, IA, MN, WA, IL.