ICU Medical, Inc.: Device Recall
Recall #Z-0154-2026 · 10/02/2025
Recall Details
- Recall Number
- Z-0154-2026
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- ICU Medical, Inc.
- Status
- Ongoing
- Date Initiated
- 10/02/2025
- Location
- San Clemente, CA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 23203
Reason for Recall
IV Gravity burette administration set burette component is missing an internal shutoff valve intended to stop fluid flow, which if used may result in delay in therapy during setup, over delivery/unrestricted flow, or air may be infused into the body.
Product Description
IV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set (w/Shut Off), MicroClave, Clamp, REF: B9213; 70" (178 cm) 150 ml Burette Set (Clave, Shut-Off) w/3-Way Stopcock, Clave, Graduated Adapter, Rotating Luer, 1 Ext, REF: B9732; 80" (203 cm) Y-Type 150 ml, Burette Set (Clave, Shut-Off), w/3-Way Stopcock, Clave, Graduated Adapter, Rotating, Luer, 1 Ext, REF: B9733; 53 cm (21") Add-On 150 ml Burette Set (Clave, Shut-Off), Vented Cap, REF: 011-C7014
Distribution Pattern
Worldwide distribution: US (nationwide) to states of: TX, SC and OUS (foreign) to countries of: Canada, Australia, and New Zealand.