ICU Medical, Inc.: Device Recall

Recall #Z-0154-2026 · 10/02/2025

Class I: Dangerous

Recall Details

Recall Number
Z-0154-2026
Classification
Class I
Product Type
Device
Recalling Firm
ICU Medical, Inc.
Status
Ongoing
Date Initiated
10/02/2025
Location
San Clemente, CA, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
23203

Reason for Recall

IV Gravity burette administration set burette component is missing an internal shutoff valve intended to stop fluid flow, which if used may result in delay in therapy during setup, over delivery/unrestricted flow, or air may be infused into the body.

Product Description

IV Administration sets. 23" (58 cm) 150 mL Burette Set (Clave, Shut Off) w/Clave, Luer Lock, REF: B33359; 23" (58 cm) 150 ml Burette Set (w/Shut Off), MicroClave, Clamp, REF: B9213; 70" (178 cm) 150 ml Burette Set (Clave, Shut-Off) w/3-Way Stopcock, Clave, Graduated Adapter, Rotating Luer, 1 Ext, REF: B9732; 80" (203 cm) Y-Type 150 ml, Burette Set (Clave, Shut-Off), w/3-Way Stopcock, Clave, Graduated Adapter, Rotating, Luer, 1 Ext, REF: B9733; 53 cm (21") Add-On 150 ml Burette Set (Clave, Shut-Off), Vented Cap, REF: 011-C7014

Distribution Pattern

Worldwide distribution: US (nationwide) to states of: TX, SC and OUS (foreign) to countries of: Canada, Australia, and New Zealand.