LEICA BIOSYSTEMS NUSSLOCH GMBH: Device Recall
Recall #Z-0086-2025 · 09/04/2024
Class II: Risk
Recall Details
- Recall Number
- Z-0086-2025
- Classification
- Class II
- Product Type
- Device
- Recalling Firm
- LEICA BIOSYSTEMS NUSSLOCH GMBH
- Status
- Ongoing
- Date Initiated
- 09/04/2024
- Location
- Nussloch, Germany
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 13 units
Reason for Recall
As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.
Product Description
Leica Biosystems Cryostat Model Number CM1510 S, Material Numbers: a) 491510C002, b) 1491510C003; for freezing and sectioning tissue samples
Distribution Pattern
US Nationwide distribution.