LEICA BIOSYSTEMS NUSSLOCH GMBH: Device Recall

Recall #Z-0085-2025 · 09/04/2024

Class II: Risk

Recall Details

Recall Number
Z-0085-2025
Classification
Class II
Product Type
Device
Recalling Firm
LEICA BIOSYSTEMS NUSSLOCH GMBH
Status
Ongoing
Date Initiated
09/04/2024
Location
Nussloch, Germany
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Reason for Recall

As part of our post market surveillance, we recognized that some customers are using flammable freezing sprays within the cryostat chamber. All flammable freezing sprays can potentially ignite and therefore can cause serious injuries. The flammability is indicated on the labelling of these sprays. The cryostat labelling, however, did not include any specific warning about usage of flammable freezing sprays prior to 2019.

Product Description

Leica Biosystems Cryostat Model Number CM1100, Material Number 14046931130, for freezing and sectioning tissue samples

Distribution Pattern

US Nationwide distribution.

Other Recalls by LEICA BIOSYSTEMS NUSSLOCH GMBH

View all recalls by LEICA BIOSYSTEMS NUSSLOCH GMBH →

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.