Boston Scientific Corporation: Device Recall
Recall #Z-0081-2026 · 08/20/2025
Class I: Dangerous
Recall Details
- Recall Number
- Z-0081-2026
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- Boston Scientific Corporation
- Status
- Ongoing
- Date Initiated
- 08/20/2025
- Location
- Saint Paul, MN, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 50,135 units
Reason for Recall
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
Product Description
ACCOLADE SR SL MRI (Model L310)
Distribution Pattern
Worldwide
Other Recalls by Boston Scientific Corporation
- Class I: Dangerous 12/19/2025
- Class I: Dangerous 12/19/2025
- Class I: Dangerous 12/19/2025
- Class I: Dangerous 12/19/2025
- Class I: Dangerous 12/19/2025
Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.