B BRAUN MEDICAL INC: Device Recall
Recall #Z-0070-2026 · 09/30/2025
Class I: Dangerous
Recall Details
- Recall Number
- Z-0070-2026
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- B BRAUN MEDICAL INC
- Status
- Ongoing
- Date Initiated
- 09/30/2025
- Location
- Allentown, PA, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1298454 units
Reason for Recall
The potential for micro-air bubbles observed in the bloodline and air-in-line alarms due to damaged arterial and venous patient connectors.
Product Description
Brand Name: Hemodialysis Bloodlines Product Name: STREAMLINE BLOODLINE SET FOR DIALOG Model/Catalog Number: SL-2010M2096 Software Version: N/A Product Description: STREAMLINE BLOODLINE SET FOR DIALOG Component: N/A
Distribution Pattern
Domestic: US Nationwide Distribution.
Other Recalls by B BRAUN MEDICAL INC
- Class II: Risk 10/29/2025
- Class II: Risk 10/29/2025
- Class II: Risk 10/29/2025
- Class II: Risk 10/29/2025
- Class II: Risk 10/29/2025
Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.