MEDLINE INDUSTRIES, LP - Northfield: Device Recall

Recall #Z-0053-2025 · 08/28/2024

Class II: Risk

Recall Details

Recall Number
Z-0053-2025
Classification
Class II
Product Type
Device
Recalling Firm
MEDLINE INDUSTRIES, LP - Northfield
Status
Ongoing
Date Initiated
08/28/2024
Location
Northfield, IL, United States
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
72 kits

Reason for Recall

CONMED Corporation (vendor) initiated a recall for select lot numbers of the Unimax Detachable Endo Pocket. The Detachable Endo Pocket is indicated for use as a receptacle for the collection and extraction of tissues, organs, and calculi during general and laparoscopic procedures. The tube inside the pouch, which detaches during the removal process, may stretch out from the pouch opening if it is not precisely fixed during the production process. Medline has identified that this product is utilized as a component within the Medline-branded convenience kits.

Product Description

Medline Sterile Laparoscopy Pack, DYNJ68187A (REF), containing the component ConMed/Unimax Detachable Endo Pocket

Distribution Pattern

US Nationwide distribution in the state of CA.

Other Recalls by MEDLINE INDUSTRIES, LP - Northfield

View all recalls by MEDLINE INDUSTRIES, LP - Northfield →

Data from FDA openFDA enforcement reports. This site is not affiliated with or endorsed by the FDA. Recall information may be updated by the FDA at any time.