Baxter Healthcare Corporation: Device Recall
Recall #Z-0046-2025 · 09/12/2024
Class I: Dangerous
Recall Details
- Recall Number
- Z-0046-2025
- Classification
- Class I
- Product Type
- Device
- Recalling Firm
- Baxter Healthcare Corporation
- Status
- Ongoing
- Date Initiated
- 09/12/2024
- Location
- Deerfield, IL, United States
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 10,957 units
Reason for Recall
The Life2000 ventilator may fail to initiate the Low Gas Pressure alarm if the pressure gas source (Life2000 compressor, oxygen cylinder or wall source) is not supplied to the ventilator before initiating therapy.
Product Description
Welch Allyn Life2000 Ventilation System, REF MS-01-0118, within exterior carton codes: 1) BT-20-0002, 2) BT-20-0002A, 3) BT-20-0002AP, 4) BT-20-0007, 5) BT-20-0011, and 6) RMS010118CP
Distribution Pattern
US nationwide distribution.